Kalibratie van styli
Deze pagina is momenteel niet beschikbaar in uw taal. U kunt met behulp van Google Translate een automatische vertaling bekijken. Wij zijn niet verantwoordelijk voor deze dienstverlening en het vertaalresultaat is niet door ons gecontroleerd.
The effective dimensions of the probe components have to be established if good results are to be obtained. These values are stored in the machine's data processor. In the case of CMMs, this may involve calibrating several styli or sensors, often in multiple head orientations. Calibration of machine tool inspection probes is generally simpler. How it worksThe position of the individual ball styli and their diameters are established using a special probe calibration programme (see the machine manufacturer's user manual). You contact a reference point with all the styli used, one after the other. The reference used is usually an extremely precise, manufactured ball with a known diameter. The exact dimensions of the ball being calibrated are input to the measuring software. If the styli are to be used for measuring separate points, the stylus is calibrated using 5-6 points on the reference ball's high points (see image).
Make sure, particularly if you are using more than one CMM, that you use the calibrated ball whose values have been input to the software. Useful linksVolgende stapNeem online contact met ons op als u meer informatie wenst of een prijsaanvraag wilt doen. U kunt ook rechtstreeks praten met uw plaatselijke Renishaw-leverancier. |